QUALITY CONTROL OF LINEAR ACCELERATOR OF NOUAKCHOTT.
- Univesity of Ibn Tofail, Kenitra (Morocco).
- Higher Institute of Nursing Professions and Health Techniques, Agadir, Morocco.
- Abstract
- Keywords
- References
- Cite This Article as
- Corresponding Author
At the National Center of Oncology of Nouakchott, the external radiotherapy treatments are performed using a linear accelerator particles of two energies 6 MV and 18 MV. This particle accelerator is originally set to deliver a dose under reference conditions. It is essential that the measured data on the particle accelerator be consistent with the data calculated by the Treatment Planning System (TPS). The purpose of this work is to perform a general quality control of this accelerator, and to compare the results obtained by two ionization chambers (PTW 0.125 cm3 and PTW 0.6 cm 3). This control is based on the measure of percentage depth dose (PDD) and beam profiles (BP) and the comparison of these measurements with the same results performed by TPS calculations for different field sizes. In general, the results measured are in good concordance with the results calculated by TPS. Regarding comparison between the two ionizations chambers, we notice that the light difference between the percentage depth doses measured by the two ionizations is related to the energy and to the field size of the incident photons beam.
- McKerracher and D. I. Thwaites; Assessment of new small-field detectors against standard-field detectors for practical stereotactic beam data acquisition; Phys. Med. Biol, 44 (1999) 2143-2160.
- Charles W. Coffey II, J. Larry Beach, Donald J. Thompson, and Marta Mendiondo; Xray beam characteristics of the Varian Clinac 6100 linear accelerator;Medical pysics, 7(1980)716-722.
- Dorin Dusciac, Jean-Marc Bordy, Josiane Daures et Valentin Blideanu; R?f?rence nationale pour la dosim?trie des rayonnements photoniques de haute ?nergie pour l??talonnage des dosim?tres et d?bitm?tres de radioprotection; Revue fran?aise de m?trologie, 42 (2016) 17-26.
- George X Ding; Energy spectra angular spread fluence profiles and dose distributions of 6 and 18 MV photon beams: results of Monte Carlo simulations for a Varian 2100EX accelerator; Phys. Med. Biol, 47 (2002) 1025-1046.
- Krithivas and S. N. Rao; A study of the characteristics of radiation contaminants within a clinically useful photon beam; Medical Physics, 12 (1985) 764-768.
- Haluk Yucel, Ibrahim Cobanbas? Asuman Kolbasi, Alptug Ozer Yuksel, and Vildan Kaya; Measurement of Photo-Neutron Dose from an 18-MV Medical Linac Using a Foil Activation Method in View of Radiation Protection of Patients; Nuclear Engineering and Technology, 48 (2016)225-235.
- James A. Purdy; Buildup/surface dose and exit dose measurements for 6MV linear accelerator; Medical Physics, 13 (1986) 259-262.
- Jean Claude Rosenwald, Laurent Bonvalet, Jocelyne Mazurier, Christine M?tayer?; Recommandations pour la mise en service et l?utilisation d?un syst?me de planification de traitement en radioth?rapie (TPS), RAPPORT SFPM N? 27,2010.
- JP Manens, I Buchheit, H Beauvais, G Gaboriaud, A Mazal, P Piret?; Dosim?trie des faisceaux de photons de faibles dimensions?; Cancer/Radiother,2 (1998) 105-114.
- -A. Johansson , J. C. Horiot , J. Van Dam , D. Lepinoy , I. Sentenac and G. Sernbo ;
- L?one Blazy-aubignac; Quality control of the treatment planning systems dose calculations in external radiation therapy using the penelope monte-carlo code; University of Toulouse III ? Paul Sabatier; 2007.
- Mawenn Le Roy; Study of national dosimetric standards for external beam radiotherapy: application to conformal irradiations; University of Nice Sophia-Antipolis; 2012.
- Mark J. Engler and Gary L. Jones; Small beam calibration by 0.6 and 0.2cm3 ionization chambers; Medical Physics, 11 (1984) 822-826.
- Miljenko Markovic, Sotirios Stathakis, Panayiotis Mavroidis, Ines-Ana Jurkovic and Nikos Papanikolaou; Characterization of a two-dimensional liquid-filled ion chamber detector array used for verification of the treatments in radiotherapy; Medical Physics, 41 (2014) 051704- 051704-15.
- Villani, K. G?rard, V. Marchesi, S. Huger, P. Fran?ois, A. No?l;?Statistical process control applied to intensity modulated radiotherapy pretreatment controlswith portal dosimetry; Cancer/Radioth?rapie,14 (2010)189?197.
- Ould Mohamed Yeslem Ahmed El Mouna, Ould Cheibetta Moussa, Choukri Abdelmajid, Ghassoun Jilali, Hakam Oum Keltoum and Semghouli Slimane; Preliminary results of quality control for linear accelerator at Oncology Center in Nouakchott; British Journal of Medicine & Medical Research, 22 (2017) 1-8.
- Sang Hyoun Choi, Kum-Bae Kim and Young Hoon Ji; Determination of Ion Recombination Correction Factors for a Liquid Ionization Chamber in Megavoltage Photon Beams; Journal of the Korean Physical Society, 66 (2015) 1439-
[Ould Mohamed Yeslem Ahmed El Mouna, Choukri Abdelmajid, Hakam Oum Keltoum and Semghouli Slimane. (2018); QUALITY CONTROL OF LINEAR ACCELERATOR OF NOUAKCHOTT. Int. J. of Adv. Res. 6 (Jan). 361-369] (ISSN 2320-5407). www.journalijar.com
Physique