COMPARATIVE STUDY OF ACCURACY OF BIOMETRIC FORMULAS OBTAINED BY APPLANATION A-SCAN BIOMETRY.
- Junior Resident, Ophthalmology, Andhra Medical College, Visakhapatnam.
- Assistant professor, Ophthalmology, Andhra Medical College, Visakhapatnam.
- Abstract
- References
- Cite This Article as
- Corresponding Author
Aim of the study: To evaluate and compare the accuracy of different intra ocular lens (IOL) power calculation formulas for predicting post operative refraction in eyes of different axial lengths using applanation A scan biometry Methods: In prospective study, conducted in GREH, Visakhapatnam with 50 eyes.Axial lengths were calculated by applanation A-scan biometry. All eyes were divided into three groups according to AL: short (<22.0 mm), average (22.0?25.0 mm), and long (>25.0 mm) eyes. IOL power calculation was done using the five Formulas SRKII, SRK/T, HOLLADAY 1, HOFFERQ, and BINKHORST 2. All patients were subjected to SICS with PCIOL implantation under local anaesthetia.The implanted power of IOL was based on SRK II formula. Final refraction was measured 6 weeks postoperatively.Post-operative spherical equivalent (SE) was calculated and compared with predicted refractive error using each biometric formula. Results: All formulas had a significantly lower MAE in comparison with BINKHORST II formula. SRK II and SRK/T had lowest MAE (0.58?0.44) for all eyes. SRK II had the lowest MAE for average (0.54?0.40) and short (0.67?0.54)eyes.SRK/T had the lowest MAE (0.63?0.48) for long eyes. BINKHORST II formula had highest MAE for all (0.69?0.48) and average (0.68?0.46) eyes. HOFFER Q formula had highest MAE (0.76?0.65) for short eyes and SRK II had highest MAE (0.89?0.64) for long eyes Conclusion:SRK II and SRK/T are accurate formulas for average eyes.SRK II and HOLLADAY I are best formula of choice in short eyes.SRK/Tis accurate IOL formula for long eyes followed by HOLLADAY I and HOFFER Q .
- O ?zcura, s. Aktas ?h. M. Sag ˘dik, m. Tetikog ˘lu (2015) comparison of the biometric formulas used for applanation a-scan ultrasound biometry, intophthalmol,feb 2016:doi 10.1007/s10792-016-0195-6.
- Hoffer kJ (1993) the hoffer q formula: a comparison of theoretic and regression formulas. J cataract refract surg 19:700?712.
- Ashish mitra, eleshjain, aloksen, shubhitripathi ,a study regarding efficacy of various intraocular lens power calculation formulas in a subset of indian myopic population,indian journal of ophthalmol,2014;62;7;826-828.
[V a n p kumari oruganti, k. Venkateswarlu, kaavyahanuraj kolalapudi and p.shyam sundar. (2019); COMPARATIVE STUDY OF ACCURACY OF BIOMETRIC FORMULAS OBTAINED BY APPLANATION A-SCAN BIOMETRY. Int. J. of Adv. Res. 7 (Feb). 362-365] (ISSN 2320-5407). www.journalijar.com