STUDY OF SOME ESSENTIAL QUALITY CONTROL TEST PARAMETERS FOR THE PERFORMANCE OF DOSE CALIBRATORS IN NUCLEAR MEDICINE PRACTICES.

- Health Physics Division, Atomic Energy Centre Dhaka, Bangladesh Atomic Energy Commission, 4 Kazi Nazrul Islam Avenue, Shahbag, Dhaka-1000, Bangladesh.
- Institute of Nuclear Medicine and Allied Sciences, Dhaka Medical College Hospital Campus, Dhaka, Bangladesh.
- Health Physics & Radioactive Waste Management Unit, Institute of Nuclear Science and Technology, Atomic Energy Research Establishment, Bangladesh Atomic Energy Commission, Ganakbari, Savar, Dhaka, Bangladesh.
- Department of Nuclear Engineering, University of Dhaka, Bangladesh.
- Abstract
- Keywords
- References
- Cite This Article as
- Corresponding Author
The aim of this study is to evaluate and compare the performance of two dose calibrators which are used in Nuclear Medicine Center by standard tests accuracy, constancy, linearity and geometry. To check the performance status of the dose calibrators the activity of radiopharmaceuticals must be accurately known before administered to the patients. Two dose calibrators of Institute Nuclear Medicine and Allied Sciences, Dhaka (Capintec, Inc.,Model no. CRC 25R and Veenstra , model no. VDC 404) performances were compared in accordance with varying geometry because most common measurement geometries utilized in clinical practice are glass vials, solid capsules and various types of plastic syringes. Vials and syringes filling volumes can be widely different, according to clinical needs. It was found that the measurement activity readings of one of these two dose calibrator, VDC-404, showed slightly deviating from accurate value because readings might be affected by the frequent change in background count. The results of the quality control (QC) tests revealed that the parameters that were traced for both dose calibrators were within the limits of the international standards, ?5%. Regular QC should cover precision, accuracy, linearity, and geometry of instrument, according to IAEA standards, e.g. (IAEA TECDOC-602 and 1599), that will guarantee to assure the proper practice of nuclear medicine diagnosis. It is also vital to fulfill radiation protection requirements as well as successful therapies and good quality image production in nuclear medicine practices.
- Calhoun, J.M.; Golas, D.B.; Harris, S.G., Effects of varying geometry on dosecalibrator response: Cobalt-57 and Technetium-99m, Journal of Nuclear Medicine, 28,1478-1483 (1987).
- Capintec, Inc. Manual stock no. 9250-0122, rev. B ? (August 2007), NJ 07446 , USA CRC-25R? Owner?s manual.
- Ceccatelli, A., Benassi, M., D?Andrea, M., De Felice, P., Fazio, A., Nocentini, S., & Strigari, L. (2007). Experimental determination of calibration settings of a commercially available radionuclide calibrator for various clinical measurement geometries and radionuclides.?Applied radiation and isotopes,?65(1), 120-125.
- IAEA (May 1991). Quality Control of Nuclear Medicine Instruments. International Atomic Energy Agency, Vienna, publication IAEA?TECDOC ? 602.
- ICRP Publication 103, 2008.Recommendations of the ICRP, Annals of the ICRP,37.
- Mo, L., Reinhard, M. I., Davies, J. B., Alexiev, D., & Baldock, C. (2006). Calibration of the Capintec CRC-712M dose calibrator for 18F.?Applied radiation and isotopes,?64(4), 485-489.
- Nichols KJ, Bacharach SL, Bergmann SR, et al. Instrumentation quality assurance and J Nucl Cardiol. 2006;13:e25?e41.
- Report of AAPM (American Association of Physicists in Medicine) Task Group 181, June 2012
- Santos, J. A. M., Carrasco, M. F., Lencart, J., & Bastos, A. L. (2009). Syringe shape and positioning relative to efficiency volume inside dose calibrators and its role in nuclear medicine quality assurance programs.?Applied Radiation and Isotopes,?67(6), 1104-1109.
- Schrader, H., Kossert, K., & Mintcheva, J. (2008). Calibration of a radionuclide calibrator system as a Bulgarian standard for activity.?Applied Radiation and Isotopes,?66(6-7), 965-971.
- Suhaib Alameen, Ahamed M. bdelfatah, Mubarak Almubarak, Assessment of Dose Calibrators Performance in Nuclear Medicine Department in Sudan, Sch. Acad. J. Pharm., June 2016; 5(6):245-250.
- Wurdiyanto, G., Marsoem, P., Candra, H., & Wijono, P. (2012). Radioactivity measurement of 18F in 16 ml vials for calibration of radionuclide calibrators.?Applied Radiation and Isotopes,?70(9), 2237-2239.
- Zanzonico, P. Routine Quality Control of Clinical Nuclear Medicine Instrumentation: A Brief Review. Journal of Nuclear Medicine. 2008; 49(7); 1114-1131
- Zeinali, H. Z. (2010). An investigation on the performance of dose calibrators in nuclear medicine centers in Iran.?Iranian Journal of Nuclear Medicine,?18(Supplement 1), 84.
[A.K.M. Mizanur Rahman, S. Reza, Mahfuza Begum, Mizanur Rahman, F. Reaz, Mahbuba Begum, R. Islam, J. Ferdous, S. Pervin, M. M. M. Siraz and S. Yeasmin. (2019); STUDY OF SOME ESSENTIAL QUALITY CONTROL TEST PARAMETERS FOR THE PERFORMANCE OF DOSE CALIBRATORS IN NUCLEAR MEDICINE PRACTICES. Int. J. of Adv. Res. 7 (Mar). 279-291] (ISSN 2320-5407). www.journalijar.com
Principal Scientific Officer, Bangladesh Atomic Energy Commission