Laboratory specialists have been demonstrated that 70% of errors occur in the pre-analytical phase which is an important service of laboratory medicine. If the occurrence of error is manmade then it can always be identified and reduced. As the hospital management needs to evaluate the quality of laboratory services by knowing the sample rejection rate. A descriptive study was conducted to investigate and reduce the rejection rate for the Biochemistry and haematology samples received in the clinical laboratory. A strategy used in the study for the source detection of Failure modes are a) Brainstorming, b) Fish-bone diagram, c) Pareto?s chart, d) Failure Mode and Effect Analysis (FMEA) is an aid to calculate Risk Priority Number (RPN) that is the product of Occurrence Index (O.I), and Detectability Index (D.I). The total number of samples (both from OP & IP-D) received in the laboratory during the study period from January 2018 to April 2019 is about 3, 81,452 samples. The total number of samples rejected by the laboratory is 182 (0.047%). The highest rates according to Pareto?s principle & FMEA analysis are found for the failure modes sample clot, insufficient sample quantity, inappropriate sample Vacutainer/tube with RPN value of 48, 36 and 30 respectively. Even though these RPN values are highest among all identified failure modes, they are less than 50 (RPN<50) which indicates that the processes involved are well controlled.

Article Analytics

References

  1. International Organization for Standardization (ISO 15189:2012): medical laboratories: particular requirements for quality and competence. Geneva, Switzerland: International Organization for Standardization; 2012
  2. Carraro P, Plebani. Errors in a stat laboratory: Types and frequency 10 years later. Clin Chem. 2007; 53: 1338-42. http://dx.doi.org/10.1373/clinchem.2007.088344.
  3. Dikmen, ZelihaGunnur et al. ?Specimen rejection in laboratory medicine: Necessary for patient safety?? Biochemiamedica vol. 25, 3 377-85, 15 oct.2015, doi 10. 11613/BM2015.037.
  4. Jacobsz, L., Zemlin, A., Roos, M., et al. (2011). Chemistry and hematology sample rejection and clinical impact in a tertiary laboratory in Cape Town.?Clinical Chemistry and Laboratory Medicine (CCLM), 49(12), pp. 2047-2050. Retrieved 29 May. 2019, from doi:10.1515/CCLM.2011.743
  5. Lao, E.G, Garcia, A.s, Figuerola M B, Moreno. E and Para ire A.H (2017) Errors of clinical Laboratory and Its Impact on Patient Safety, Open Journal of Social Science, 5,243-253.
  6. Plebani M. Quality Indicators to Detect Preanalytical Errors in Laboratory Testing. ClinBiochem Rev 2012; 33:85-8.
  7. Nikunj Modi, Tejas Shah; Application of Six sigma test in clinical biochemistry laboratory, International journal of research medicine,Int J Res Med. 2017; 6(2); 75-78
  8. Laboratory Quality Control Based on Risk Management; Approved Guideline. CLSI document. EP23-A. Wayne, PA: Clinical and Laboratory Standards Institute, 2011.
  9. Plebani M, Lippi G. To err is human. To misdiagnosis might be deadly. ClinBiochem 2010; 43:1-3.
  10. LlopisMA, Trujillo G, Llovet MI, Tarres E, Ibarz M, Biosca C et al. Quality indicators and specifications for key analytical-extranalytical processes in the clinical laboratory. Five years experience using the Six Sigma concept. ClinChem Lab Med 2011; 49:463-70. http://dx.doi.org/10.1515/CCLM.2011.067.
  11. Gim?nez-Mar?n, Angeles et al. ?Pre-analytical errors management in the clinical laboratory.? Biochemia media vol.24, (2): 248-57. doi:10.11613/BM.2014.027.

How to Cite This Article

Golla Rachana and Manjunatha R. (2019); A STUDY ON IDENTIFYING THE PRE-ANALYTICAL PHASE ERRORS LEADING TO BIOCHEMISTRY AND HEMATOLOGY SAMPLE REJECTION BY USING FMEA & PARETOS PRINCIPLE., Int. J. of Adv. Res., 7 (09), 811-820, ISSN 2320-5407. DOI: https://doi.org/10.21474/IJAR01/9731

Corresponding Author

Dr. Manjunatha R
IIHMR, Bangalore