CORRELATION OF INTRAOCULAR PRESSURE MEASURED BY SCHIOTZ INDENTATION TONOMETER AND GOLDMANN APPLANATION TONOMETER

  • Junior Resident, Department of Ophthalmology, DBVP Rural Medical College, Loni.
  • Professor, Department of Ophthalmology, DBVP Rural Medical College, Loni.
18 Downloads 80 Views
Crossref

Abstract

Aim: To study and correlate the Intraocular pressure measured by Schiotz indentation tonometer and Goldmann applanation tonometer.

Materials and Methods: Cross-sectional Analytical study was done in 200 eyes of 100 purposively selected patients and were subjected to two methods of tonometry - Goldmann applanation tonometry and Schiotz indentation tonometry in a period of 10 months from January 2024 to October 2024. Correlation between the two values was then observed and the agreement between the two was plotted on Bland Altman plot. Results- There was positive correlation found between the IOP measured by Schiotz indentation tonometer and Goldmann Applanation tonometer and there was no systematic difference found between the two on Bland Altman plot.

Conclusion: In conclusion, IOP readings using Schiotz indentation tonometer do correlate with Goldmann applanation tonometer. Hence, as being portable and relatively cost effective, Schiotz tonometer can be used in rural settings for baseline IOP measurement.

How to Cite This Article

Shinde Revati Pravin and Waman Chavan (2025); CORRELATION OF INTRAOCULAR PRESSURE MEASURED BY SCHIOTZ INDENTATION TONOMETER AND GOLDMANN APPLANATION TONOMETER, International Journal of Advanced Research (IJAR), 13 (01), 1324-1328, ISSN 2320-5407. DOI: https://doi.org/10.21474/IJAR01/20321

Corresponding Author

Dr. Shinde Revati Pravin
Junior Resident
India

Article Analytics