AN INTEGRATIVE DESIGN APPROACH FOR ADAPTING OPEN-SOURCE MEDICAL DEVICES TO RESOURCE-LIMITED SETTINGS: DESIGN, LOCAL MANUFACTURE AND TECHNICAL EVALUATION OF THE LAAFI-RESPIR VENTILATOR IN BURKINA FASO
- Centre National de la Recherche Scientifique et Technologique (CNRST), Institut de Recherche en Sciences Appliquees et Technologies (IRSAT), Ouagadougou, Burkina Faso.
- Universite Norbert ZONGO (UNZ), Laboratoire de Chimie Analytique, de Physique Spatiale et Energetique (L CAPSE), Doctoral School of Science and Technology, Koudougou, Burkina Faso.
- Centre National de la Recherche Scientifique et Technologique (CNRST), Institut de Recherche en Sciences de la Sante (IRSS), Ouagadougou, Burkina Faso.
1 Downloads
11 Views
Abstract
In low-income countries, many medical devices sit unused or out of service because local manufacturing, maintenance and energy conditions were not considered at the design stage. The open-source ventilators produced during the COVID - 19 pandemic rely on components that are rarely available in Sahelian countries. This study proposes and tests a design approach that adapts an existing open-source solution to the local context. A six-step approach, in which contextualization is carried out during functional analysis through the network of local manufacturing and maintenance actors, was applied to LAAFI-RESPIR, a bag-valve mechanical ventilator built at IRSAT (Burkina Faso). Five open-source projects were scored on five criteria. Bench tests measured the volume delivered by five cams over ten cycles each and the respiratory rate over controller settings of 40 to 80%. Bias, repeatability and equivalence with the MHRA tolerance (two one-sided tests) were analyzed, and the prototype was compared with the MHRA specification. OxyGEN was selected (13 of 15 points). Functional analysis added eleven constraint functions covering maintenance, local manufacturing, energy, safety and mobility.
Keywords
How to Cite This Article
Frederic Bationo et,al (2026); AN INTEGRATIVE DESIGN APPROACH FOR ADAPTING OPEN-SOURCE MEDICAL DEVICES TO RESOURCE-LIMITED SETTINGS: DESIGN, LOCAL MANUFACTURE AND TECHNICAL EVALUATION OF THE LAAFI-RESPIR VENTILATOR IN BURKINA FASO, International Journal of Advanced Research (IJAR), 14 (11), 562-580, ISSN 2320-5407.